4.7 Article

Feasibility and Safety of Tethered Capsule Endomicroscopy in Patients With Barrett's Esophagus in a Multi-Center Study

期刊

CLINICAL GASTROENTEROLOGY AND HEPATOLOGY
卷 20, 期 4, 页码 756-+

出版社

ELSEVIER SCIENCE INC
DOI: 10.1016/j.cgh.2021.02.008

关键词

Barrett's Esophagus; Optical Coherence Tomography; Tethered Capsule Endomicroscopy

资金

  1. NIH [R01CA184102]
  2. Remondi Family Foundation
  3. Hazard Family Foundation
  4. MGH Research Scholars Program

向作者/读者索取更多资源

TCE can be used at multiple sites, administered to unsedated patients by non-expert physicians or nurses, and rapidly and safely evaluate the microscopic structure of the esophagus.
BACKGROUND & AIMS: Tethered capsule endomicroscopy (TCE) involves swallowing a small tethered pill that implements optical coherence tomography (OCT) imaging, procuring high resolution images of the whole esophagus. Here, we demonstrate and evaluate the feasibility and safety of TCE and a portable OCT imaging system in patients with Barrett's esophagus (BE) in a multi-center (5-site) clinical study. METHODS: Untreated patients with BE as per endoscopic biopsy diagnosis were eligible to participate in the study. TCE procedures were performed in unsedated patients by either doctors or nurses. After the capsule was swallowed, the device continuously obtained 10-um-resolution cross-sectional images as it traversed the esophagus. Following imaging, the device was withdrawn through mouth, and disinfected for subsequent reuse. BE lengths were compared to endoscopy findings when available. OCT-TCE images were compared to volumetric laser endomicroscopy (VLE) images from a patient who had undergone VLE on the same day as TCE. RESULTS: 147 patients with BE were enrolled across all sites. 116 swallowed the capsule (79%), 95/114 (83.3%) men and 21/33 (63.6%) women (P = .01). High-quality OCT images were obtained in 104/111 swallowers (93.7%) who completed the procedure. The average imaging duration was 5.55 +/- 1.92 minutes. The mean length of esophagus imaged per patient was 21.69 +/- 5.90 cm. A blinded comparison of maximum extent of BE measured by OCT-TCE and EGD showed a strong correlation (r = 0.77-0.79). OCT-TCE images were of similar quality to those obtained by OCT-VLE. CONCLUSIONS: The capabilities of TCE to be used across multiple sites, be administered to unsedated patients by either physicians or nurses who are not expert in OCT-TCE, and to rapidly and safely evaluate the microscopic structure of the esophagus make it an emerging tool for screening and surveillance of BE patients.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据