Comparison between the portable pressure measuring device and PicoPress® for garment pressure measurement on hypertrophic burn scar during compression therapy

期刊

BURNS
卷 47, 期 7, 页码 1621-1626

出版社

ELSEVIER SCI LTD
DOI: 10.1016/j.burns.2021.01.018

关键词

Pressure measuring device; Burn; Hypertrophic scar; Silicon piezoresistive; pressure sensor

资金

  1. Ministry of Education [NRF-2018R1C1B5044080]
  2. Soon-chunhyang University Research Fund

向作者/读者索取更多资源

The study developed a novel portable pressure measuring device for accurate pressure measurements at the hypertrophic scar-pressure garment interface. Results showed that the test-retest reliability of the novel device was close to the acceptable level for clinical use, and the pressure values obtained from the device were consistent with those from the traditional PicoPress(R) measurements.
Purpose: The current standard treatment for hypertrophic scars following burn injury is pressure garment therapy. The experimenters developed the novel portable pressure measuring device using silicon piezoresistive sensors. As PicoPress (R) is the most accurate (i.e., lowest variation and error) manometric sensor for pressure measurement, we sought to compare and examine the accuracy of the novel device regarding in vitro pressure measurements at the hypertrophic scar-pressure garment interface. Methods: The novel device was designed to operate in non-corrosive media, such as air. The device can use up to six pressure sensing points and was developed to adjust the number of pressure sensors according to the size of the scar. Pressure measurements were acquired through a readout circuit consisting of an analog-to-digital converter, a microprocessor, and a Bluetooth transmission module for wireless data transmission to an external device. All signals were converted into mean pressure expressed in millimeters of mercury (mmHg). The mean pressure values measured by the sensors were compared to those obtained from PicoPress (R). 55 garment pressures recordings were obtained from the sensors over this study conducted in 2018-February 2020. We then analyzed the test-retest reliability using the intraclass correlation coefficients (ICC). PicoPress (R) was also employed in the same pressure garments for obtaining similar measurements. A two way random effects model ICC with 95% confidence intervals was used to compare the mean pressure values obtained from the silicon piezoresistive sensors to the PicoPress (R) measurements. Results: The test-retest reliability of the pressure sensors was close to the acceptable level for clinical use regarding stationary interface pressure measurement (ICC = 0.99,95% CI 0.990-0.997). The mean pressure obtained from the silicon piezoresistive pressure sensors showed an accordance with the measurements from PicoPress (R) (ICC = 0.97, 95% CI 0.947-0.985). Conclusion: The novel device may present a viable alternative to PicoPress (R) for garment pressure measurements. In addition, the novel device improves adaptability to the hypertrophic scar shape and size. Complementary characteristics such as wireless transmission to an external device may allow burn patients to continuously wear the device for real-time measurements during pressure garment therapy, thus improving existing devices including PicoPress (R). (C) 2021 Elsevier Ltd and ISBI. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据