4.4 Article

Assessing the microscale heterogeneity in Standard Reference Material 4600 Surrogate Post-detonation Urban Debris

期刊

APPLIED RADIATION AND ISOTOPES
卷 172, 期 -, 页码 -

出版社

PERGAMON-ELSEVIER SCIENCE LTD
DOI: 10.1016/j.apradiso.2021.109651

关键词

Microheterogeneity; SRM; PCA; Surrogate debris; X-ray fluorescence

资金

  1. FBI Laboratory Division
  2. NIST Radioactivity Group

向作者/读者索取更多资源

Nondestructive microbeam X-ray fluorescence (mu XRF) spectrometry was used to investigate elemental microheterogeneity in a nuclear forensics reference material, showing most elements are homogeneous except for zinc. To minimize measurement uncertainty from elemental microheterogeneity, a minimum sample mass of 24 mg is recommended for analysis.
Nondestructive microbeam X-ray fluorescence (mu XRF) spectrometry has been used to investigate the elemental microheterogeneity in a nuclear forensics reference material (RM), NIST SRM 4600 Surrogate Post-detonation Urban Debris. Using a principal component analysis (PCA) model, results indicate the majority of elements appear homogeneous; however, zinc (Zn) exhibits microscale heterogeneity for this SRM. To minimize contributions to the measurement uncertainty from elemental microheterogeneity, a minimum sample mass of 24 mg is recommended for analysis.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据