4.5 Article

Comparison of Sampling Methods for International Normalized Ratio Monitoring in Haemodialysis Patients (INRHaemo Study)

期刊

AMERICAN JOURNAL OF NEPHROLOGY
卷 52, 期 1, 页码 17-25

出版社

KARGER
DOI: 10.1159/000513094

关键词

Haemodialysis; Warfarin; International normalized ratio; Point-of-care; Sampling

向作者/读者索取更多资源

Good agreement and correlation were found between peripheral blood, haemodialysis access/circuit, and point-of-care INR values in dialysis patients on warfarin. Point-of-care INR was less reliable at higher values, and dosing decisions differed from peripheral INR in a quarter of cases.
Background: Haemorrhagic and thrombotic complications are common in dialysis patients on warfarin; thus, accurate international normalized ratio (INR) monitoring is critical. For expediency and patient comfort, blood sampling from the haemodialysis access or circuit is commonly performed. Point-of-care (POC) INR machines allow both peripheral vein preservation and rapid results, yet are not validated in the haemodialysis population. Methods: A prospective cohort study in haemodialysis patients taking warfarin was undertaken. Three paired samples were drawn over a single session: peripheral blood INR, POC INR, and dialysis INR. Agreement using Bland-Altman analysis and correlation coefficients between the peripheral blood INR, haemodialysis INR, and POC INR were calculated. Inappropriate dosing decisions based on haemodialysis or POC INR were quantified. Results: Amongst 34 patients, agreement between the dialysis INR and peripheral blood INR was high, with the haemodialysis INR differing from the peripheral INR by <+/- 0.2, 85.2% of the time. Correlation between the 2 methods was high (r = 0.914; p < 0.001). POC INR differed from peripheral INR values by <+/- 0.2, 67.6% of the time, with less agreement at higher INR values. Dosing decisions were incongruent between the dialysis and peripheral INR in 6%, whilst the POC and peripheral INR disagreed in 26%. Conclusions: There was good agreement and correlation between the peripheral blood, haemodialysis access/circuit, and POC INR values. POC INR was less reliable at higher values, and dosing decisions differed from the peripheral INR in a quarter of cases.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据