A Computer Simulation Research of Two Types of Cardiac Physiological Pacing

期刊

APPLIED SCIENCES-BASEL
卷 11, 期 1, 页码 -

出版社

MDPI
DOI: 10.3390/app11010449

关键词

cardiac modeling; computer simulation; is-bundle pacing; left bundle branch pacing; left bundle branch block; atrioventricular block

资金

  1. Shanghai Municipal Science and Economic and Informatization Commission Project [GYQJ-2018-2-05]
  2. National Natural Science Foundation of China [1171009, 61801123]
  3. China Postdoctoral Science Foundation [2019M651367]
  4. Shanghai Municipal Science and Technology Major Project [2017SHZDZX01, 16441907900]

向作者/读者索取更多资源

This study utilized cardiac modeling and simulation to investigate physiological pacing in clinical settings. By constructing a multiscale rabbit ventricular electrophysiological model, the effects of HBP and LBBP pacing on correcting LBBB were explored, with an emphasis on identifying the optimal pacing timing for improved results.
This manuscript adopted the cardiac modeling and simulation method to study the problems of physiological pacing in clinical application. A multiscale rabbit ventricular electrophysiological model was constructed. We simulated His-bundle pacing (HBP) treatment for left bundle branch block (LBBB) and atrioventricular block (AVB), and left bundle branch pacing (LBBP) treatment for LBBB by setting various moments of the stimulus. The synthetic ECGs and detailed electrical activities were analyzed. Our electrophysiological model accurately simulated the normal state, HBP, and LBBP. The synthetic ECG showed that QRS duration was narrowed by 30% after HBP correction for LBBB. For LBBB correction with LBBP, the synthetic ECGs of LBBP starting before 30 ms (if the end of atrial excitation is set as 0 ms) presented right bundle branch block (RBBB), and those of LBBP starting at 30-38 ms were synchronous, while those of LBBP starting after 42 ms possessed LBBB morphologies. The best pacing results were obtained when LBBP started at 34 ms. This manuscript verified the feasibility of the constructed ventricular model, and studied the physiological pacing mechanism. The results showed that HBP realized correction for AVB and high LBBB. The performance of LBBP can be improved by applying the stimulus within a specific period of time (0-8 ms) after atrial excitation.