Variation in sag values in daily disposable, reusable and toric soft contact lenses

期刊

CONTACT LENS & ANTERIOR EYE
卷 44, 期 6, 页码 -

出版社

ELSEVIER
DOI: 10.1016/j.clae.2020.11.006

关键词

CL-SAG values; Daily disposables; Reusable lenses; Toric lens designs; Soft lens base curve; Diameter

资金

  1. Pacific University College of Optometry team

向作者/读者索取更多资源

Significant differences in CL-SAG values were found between different types of lenses, even within lenses with the same BC. Lens diameter and BC together influence CL-SAG values, indicating that lens substitution based solely on BC could result in substantial differences. Lens designs offering two BCs showed a narrower band in absolute terms in CL-SAG, and switching from a spherical to a toric lens of the same manufacturer could result in a significant change in CL-SAG.
Purpose: While knowledge of the ocular surface sagittal height (OC-SAG) is increasing with newer instrumentation, understanding of soft lens sag values (CL-SAG) in daily clinical practice is limited. This study aimed to report the differences in CL-SAG of a large collection of daily disposable (DD), reusable (2-week and 4-week replacement) and toric lens designs. Methods: 12 different spherical DD and 15 reusable lens types were analysed (labelled power -3.00D); 17 different toric lens types analysed (6 DD and 11 reusable) in - 3.00D =C 0.75 x 180. Posterior CL-SAG was measured directly using a spectral domain OCT-based instrument (Optimec is830). Results: The range in CL-SAG for DDs was 3,398-3,912 microns, for reusable lenses 3,372-4,004 microns and for toric lenses 3,536-4,157 microns. Lens diameter and BC together influence CL-SAG values. For lenses with the same labelled base curve (BC), the maximum difference in the DD group was 264 microns, in the reusable group 418 microns and in the toric group 345 microns. Three lens designs within the DD group and five in the reusable group offer two BCs, with an average variation (+/- SD) in CL-SAG of 276 +/- 39 microns; one toric lens with two BCs showed a 377-micron difference. Conclusions: Marked differences in CL-SAG values were found in this study, also between lenses of the same BC. While ECPs should be aware that diameter and BC together influence CL-SAG, this also indicates that lens substitution based on BC alone could result in substantial differences in CL-SAG, with a potential impact on lens fit. Lens designs offering two BCs showed a somewhat narrow band in absolute terms in CL-SAG, and switching from a spherical to a toric lens of the same lens manufacturer could result in a significant change in CL-SAG.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据