4.1 Article

O-Arm Stereotactic Imaging in Deep Brain Stimulation Surgery Workflow: A Utility and Cost-Effectiveness Analysis

期刊

STEREOTACTIC AND FUNCTIONAL NEUROSURGERY
卷 99, 期 2, 页码 93-106

出版社

KARGER
DOI: 10.1159/000510344

关键词

O-arm; Deep brain stimulation; Movement disorders; Stereotactic imaging

向作者/读者索取更多资源

The use of the Medtronic O-arm device in DBS surgery significantly reduced the duration of image acquisition, exposure to ionizing radiation, and costs when compared with standard stereotactic MRI or CT, without compromising accuracy.
Introduction: Deep brain stimulation (DBS) surgery is an established treatment for movement disorders. Advances in neuroimaging techniques have resulted in improved targeting accuracy that may improve clinical outcomes. This study aimed to evaluate the safety and feasibility of using the Medtronic O-arm device for the acquisition of intraoperative stereotactic imaging, targeting, and localization of DBS electrodes compared with standard stereotactic MRI or computed tomography (CT). Methods: Patients were recruited prospectively into the study. Routine frame-based stereotactic DBS surgery was performed. Intraoperative imaging was used to facilitate and verify the accurate placement of the intracranial electrodes. The acquisition of coordinates and verification of the position of the electrodes using the O-arm were evaluated and compared with conventional stereotactic MRI or CT. Additionally, a systematic review of the literature on the use of intraoperative imaging in DBS surgery was performed. Results: Eighty patients were included. The indications for DBS surgery were dystonia, Parkinson's disease, essential tremor, and epilepsy. The globus pallidus internus was the most commonly targeted region (43.7%), followed by the subthalamic nucleus (35%). Stereotactic O-arm imaging reduced the overall surgical time by 68 min, reduced the length of time of acquisition of stereotactic images by 77%, reduced patient exposure to ionizing radiation by 24.2%, significantly reduced operating room (OR) costs per procedure by 31%, and increased the OR and neuroradiology suite availability. Conclusions: The use of the O-arm in DBS surgery workflow significantly reduced the duration of image acquisition, the exposure to ionizing radiation, and costs when compared with standard stereotactic MRI or CT, without reducing accuracy.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.1
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据