4.6 Article

Calibrations of Low-Cost Air Pollution Monitoring Sensors for CO, NO2, O3, and SO2

期刊

SENSORS
卷 21, 期 1, 页码 -

出版社

MDPI
DOI: 10.3390/s21010256

关键词

low-cost gas sensors; electrochemical air quality sensors; field evaluation; single and multiple linear regression; random forest; LSTMs; environmental factors

资金

  1. National Key R&D Program of China [2017YFB0504000]
  2. National Natural Science Foundation of China [41705113, 41877312]

向作者/读者索取更多资源

This study evaluated the performance of four electrochemical air quality sensors and different calibration models in Beijing over a 12-month period. The results showed that LSTM models outperformed random forest and linear methods in improving the accuracy of pollutant measurements, especially for O-3 and NO2.
Pollutant gases, such as CO, NO2, O-3, and SO2 affect human health, and low-cost sensors are an important complement to regulatory-grade instruments in pollutant monitoring. Previous studies focused on one or several species, while comprehensive assessments of multiple sensors remain limited. We conducted a 12-month field evaluation of four Alphasense sensors in Beijing and used single linear regression (SLR), multiple linear regression (MLR), random forest regressor (RFR), and neural network (long short-term memory (LSTM)) methods to calibrate and validate the measurements with nearby reference measurements from national monitoring stations. For performances, CO > O-3 > NO2 > SO2 for the coefficient of determination (R-2) and root mean square error (RMSE). The MLR did not increase the R-2 after considering the temperature and relative humidity influences compared with the SLR (with R-2 remaining at approximately 0.6 for O-3 and 0.4 for NO2). However, the RFR and LSTM models significantly increased the O-3, NO2, and SO2 performances, with the R-2 increasing from 0.3-0.5 to >0.7 for O-3 and NO2, and the RMSE decreasing from 20.4 to 13.2 ppb for NO2. For the SLR, there were relatively larger biases, while the LSTMs maintained a close mean relative bias of approximately zero (e.g., <5% for O-3 and NO2), indicating that these sensors combined with the LSTMs are suitable for hot spot detection. We highlight that the performance of LSTM is better than that of random forest and linear methods. This study assessed four electrochemical air quality sensors and different calibration models, and the methodology and results can benefit assessments of other low-cost sensors.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据