4.6 Article

Estimation of the Respiratory Rate from Localised ECG at Different Auscultation Sites

期刊

SENSORS
卷 21, 期 1, 页码 -

出版社

MDPI
DOI: 10.3390/s21010078

关键词

respiratory rate (RR); Electrocardiogram (ECG); ECG derived respiration (EDR); auscultation sites

资金

  1. King's China Scholarship Council

向作者/读者索取更多资源

This study investigated the accuracy of ECG-derived respiratory rate, finding that electrode placement does not impact estimation, baseline wander and amplitude modulation algorithms perform best, and frequency domain features are helpful for accurate estimation of respiratory rate.
The respiratory rate (RR) is a vital physiological parameter in prediagnosis and daily monitoring. It can be obtained indirectly from Electrocardiogram (ECG) signals using ECG-derived respiration (EDR) techniques. As part of the study in designing an early cardiac screening system, this work aimed to study whether the accuracy of ECG derived RR depends on the auscultation sites. Experiments were conducted on 12 healthy subjects to obtain simultaneous ECG (at auscultation sites and Lead I as reference) and respiration signals from a microphone close to the nostril. Four EDR algorithms were tested on the data to estimate RR in both the time and frequency domain. Results reveal that: (1) The location of the ECG electrodes between auscultation sites does not impact the estimation of RR, (2) baseline wander and amplitude modulation algorithms outperformed the frequency modulation and band-pass filter algorithms, (3) using frequency domain features to estimate RR can provide more accurate RR except when using the band-pass filter algorithm. These results pave the way for ECG-based RR estimation in miniaturised integrated cardiac screening device.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据