Is load control necessary to produce physiological AP displacement and axial rotation in wear testing of TAR?

期刊

JOURNAL OF ORTHOPAEDIC RESEARCH
卷 39, 期 4, 页码 797-805

出版社

WILEY
DOI: 10.1002/jor.24931

关键词

biomechanics; foot and ankle; implant wear

资金

  1. J. Vernon Luck, Sr., M.D. Orthopaedic Research Center at the Orthopaedic Institute for Children, UCLA

向作者/读者索取更多资源

The study conducted load-controlled testing on cadaveric specimens to evaluate total ankle replacements, revealing variations in anterior-posterior displacement and axial rotation among specimens but without exceeding the specified limits by ISO standards.
The International Standard Organization, ISO 22622, specifies two options for joint wear simulator evaluation of total ankle replacements (TARs): load-controlled and displacement-controlled. In the present study, the load-controlled testing parameters were applied to cadaveric specimens to quantify and compare the observed sagittal translations and axial rotations to those specified under the displacement-controlled option. Twelve cadaveric specimens were stripped of extraneous tissues, keeping surrounding ankle ligaments. A halo was used to produce plantarflexion and dorsiflexion of the talus through two screws, while a baseplate resisted axial loads. The axial force and torque were applied to the tibia and fibula under force and torque feedback control. An anterior-posterior force was applied to the tibia. Plantarflexion-dorsiflexion were applied using rotation control. To protect the cadaveric specimens, loads were applied at 50% of the specified load profile while plantarflexion-dorsiflexion rotation was applied as specified. There was variation among specimens in magnitudes of anterior-posterior displacement with peaks ranging from 3.3 mm posteriorly to 3.0 mm anteriorly. Likewise, there was variation among specimens in magnitude of axial rotation, with peaks ranging from 11 degrees external rotation to 4.5 degrees internal rotation. However, the mean magnitudes of AP displacement and axial rotation did not exceed those specified by ISO 22622.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据