4.7 Article

Fully automated detection and differentiation of pandemic and endemic coronaviruses (NL63, 229E, HKU1, OC43 and SARS-CoV-2) on the hologic panther fusion

期刊

JOURNAL OF MEDICAL VIROLOGY
卷 93, 期 7, 页码 4438-4445

出版社

WILEY
DOI: 10.1002/jmv.26749

关键词

coronavirus; epidemiology; pandemics; SARS coronavirus; virus classification

类别

向作者/读者索取更多资源

The Hologic Panther Fusion platform offers fully automated diagnostics for respiratory viruses including SARS-CoV-2, but not for endemic human coronaviruses. The laboratory developed test and CE marked Aptima SARS-CoV-2 TMA showed high concordance in identifying hCoV and SARS-CoV-2, expanding the diagnostic spectrum for coronaviruses globally.
The hologic panther fusion (PF) platform provides fully automated CE marked diagnostics for respiratory viruses, including the recently discovered severe acute respiratory syndrome-coronavirus 2 (SARS-CoV-2) by a transcription mediated amplification (TMA) assay, but not for the endemic human coronaviruses (hCoV). Therefore, a laboratory developed test (LDT) comprising a multiplexed reverse transcription polymerase chain reaction (RT-PCR) protocol that detects and differentiates the four hCoV NL63, 229E, HKU1, and OC43 was adapted on the PF. The novel CE marked Aptima SARS-CoV-2 TMA and the LDT for hCoV were validated with 321 diagnostic specimens from the upper and lower respiratory tract in comparison to two SARS-CoV-2 RT-PCRs (PF E-gene RT-PCR and genesig RT-PCR, 157 specimens) or the R-GENE hCoV/hParaFlu RT-PCR (164 specimens), respectively. For the endemic hCoV, results were 96.3% concordant with two specimens discordantly positive in the PF and four specimens discordantly positive in the R-GENE assay. All discordantly positive samples had Ct values between 33 and 39. The PF hCoV LDT identified 23 hCoV positive specimens as NL63, 15 as 229E, 15 as HKU1, and 25 as OC43. The Aptima SARS-CoV-2 TMA gave 99.4% concordant results compared to the consensus results with a single specimen discordantly positive. Moreover, 36 samples from proficiency testing panels were detected and typed correctly by both novel methods. In conclusion, the SARS-CoV-2 TMA and the LDT for hCoV enhanced the diagnostic spectrum of the PF for all coronaviruses circulating globally for a multitude of diagnostic materials from the upper and lower respiratory tract.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据