4.3 Article

Real-time monitoring of a positron beam using a microchannel plate in single-particle mode

期刊

JOURNAL OF INSTRUMENTATION
卷 15, 期 11, 页码 -

出版社

IOP PUBLISHING LTD
DOI: 10.1088/1748-0221/15/11/P11030

关键词

Accelerator modelling and simulations (multi-particle dynamics; single-particle dynamics); Beam-line instrumentation (beam position and profile monitors; beam-intensity monitors; bunch length monitors); Particle detectors; Real-time monitoring

向作者/读者索取更多资源

This work presents a real-time diagnostic method to monitor and characterize a positron beam by using a microchannel plate (MCP), a phosphor screen and a CCD camera. We propose a digitizing method based on single-particle detection. This approach offers the possibility to quantify the particles per unit time that interact with the MCP. This type of detector is based on a consolidated technology and its efficiency has been demonstrated for nuclear radiation and particles in many experiments. This work also presents the detection efficiency as a function of the positron kinetic energy in the range from 50 eV to 17 keV, essential for a complete characterization of the particle beam. The method allows the characterization of a positron beam in real time with a spatial resolution of the order of tens of microns.