3-(4-Carboxybenzoyl)quinoline-2-carboxaldehyde labeling for direct analysis of amino acids in plasma is not suitable for simultaneous quantification of tryptophan, tyrosine, valine, and isoleucine by CE/fluorescence

期刊

ELECTROPHORESIS
卷 42, 期 9-10, 页码 1108-1114

出版社

WILEY
DOI: 10.1002/elps.202000263

关键词

Amino acid; Biomolecules; Capillary electrophoresis; Conductivity; Fluorescence

向作者/读者索取更多资源

The study utilized capillary electrophoresis coupled to LED-induced fluorescence detection and CBQCA labeling for the analysis of amino acids such as tryptophan, facing challenges in detection and quantification, especially for Trp. Another approach, CE/capacitively coupled contactless conductivity detection, showed better performance for quantification of Trp, Tyr, Val, and Ile, with a LOD of 2 μM for Trp.
Capillary electrophoresis coupled to LED-induced fluorescence detection is a robust and sensitive technique used for amino acids (AA) analysis in biological media, after labeling with 3-(4-carboxybenzoyl)quinoline-2-carboxaldehyde (CBQCA). We wanted to quantitate in plasma tryptophan (Trp), tyrosine (Tyr), valine (Val), and isoleucine (Ile). Among the different labeled AA-CBQCA, Trp has the lowest fluorescence yield, which makes its detection and quantification very difficult in biological samples such as plasma. We tried to improve Trp analysis by CE/LED-induced fluorescence detection to its maximal sensitivity by using large volume sample stacking as a preconcentration step in our analytical protocol. At pH 9.5, this step caused a drop in resolution during the separation of the four AAs and it was therefore necessary to work at pH 10. We have found that Tyr, Val, Ile, and Trp are detected and well separated from the other AAs, but Trp cannot be quantified in plasma samples, mainly because of the low fluorescence yield of the Trp-CBQCA derivative. The recorded LOD is 0.18 mu M for Trp-CBQCA in standard solution with a resolution between Trp and Tyr of 1.2, while the LOD is 6 mu M in plasma with the same resolution. Trp, Tyr, Val, and Ile are, however, efficiently quantified when using a 3 M acetic acid electrolyte and CE associated with capacitively coupled contactless conductivity detection, which also has the advantage of not requiring derivatization or large volume sample stacking. This article demonstrates, for the CE user, that quantitative analysis of these four AA in mouse plasma can be performed by CE-fluorescence after CBQCA labeling, with the exception of Trp. It can be advantageously replaced by CE/capacitively coupled contactless conductivity detection, the only efficient one for Trp, Tyr, Val, and Ile quantification. In this case, the LOD for Trp is 2 mu M. The four AAs are separated with resolution with neighbors above 1.5.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据