Inaccurate Glucose Sensor Values After Hydroxyurea Administration

期刊

DIABETES TECHNOLOGY & THERAPEUTICS
卷 23, 期 6, 页码 443-451

出版社

MARY ANN LIEBERT, INC
DOI: 10.1089/dia.2020.0490

关键词

TPIAT; Continuous glucose monitoring; Pediatrics; Diabetes; Pharmacology

向作者/读者索取更多资源

The study evaluated the impact of hydroxyurea (HU) administration on continuous glucose monitoring (CGM) systems in patients, finding a significant rise in sensor readings from 0.5-6 hours after administration. It is recommended that patients rely on glucometer testing during this period.
Objective: To assess the degree, duration, mean absolute relative difference (MARD), and error analysis of discrepant values per continuous glucose monitoring (CGM) systems after hydroxyurea (HU) administration. Research Design and Methods: Inpatient glucometer and CGM data from 16 total pancreatectomy/islet autotransplantation patients using Dexcom Professional G4 and 12 patients using Dexcom G6 were analyzed after daily dosing with HU. Timing of HU dosing and median of 9.5 days of sensor and glucometer values were assessed per patient. Results: A large positive elevation of sensor readings was identified after HU dosing. The greatest discrepancy between glucometer and sensor readings occurred 0.5-2 h after HU administration [G4 (mean 3.0 mmol/L, median 2.4 mmol/L, MARD 55%), G6 (mean 4.2 mmol/L, median 4.6 mmol/L, MARD 91%)]. The discrepancy was <1.1 mmol/L, mean (-0.5 mmol/L) and median (-0.5 mmol/L), MARD 14% (G4) and <1.1 mmol/L, mean (0.3 mmol/L) and median (0.3 mmol/L), MARD 17% (G6), by 6 h after administration. Error analysis with the G6 system found 94% of pairs in clinically acceptable range by 6-9 h after HU administration. Aspirin, also given once daily, did not result in glucose value discrepancy with the G6 system but variability was observed with the G4 system. Conclusions: There was marked elevation of sensor glucose readings compared with glucometer values [up to 13.9 mmol/L (G4), 13 mmol/L (G6)] from 0.5 to 6 h after HU administration. It is important to counsel a patient using a Dexcom CGM system and HU therapy on this finding and to advise reliance on glucometer testing for accurate glucose assessment up to 6-9 h after HU administration.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据