Closed-Loop Oxygen Control Using a Novel Nasal High-Flow Device: A Randomized Crossover Trial

期刊

RESPIRATORY CARE
卷 66, 期 3, 页码 416-424

出版社

DAEDALUS ENTERPRISES INC
DOI: 10.4187/respcare.08087

关键词

nasal high-flow; oxygen; closed-loop; automatic titration

资金

  1. Fisher and Paykel Healthcare
  2. Health Research Council of New Zealand as an Independent Research Organisation grant

向作者/读者索取更多资源

The study demonstrates that the novel nasal high-flow device with closed-loop control can respond to changes in S-pO2 outside a target saturation range in a model of exercise-induced desaturation and subsequent recovery.
BACKGROUND: Oxygen administration is recommended for patients with hypoxemia to achieve a target S(pO2 )range. Strategies to achieve this in clinical practice are suboptimal. We investigated automatic oxygen titration using a novel nasal high-flow device with closed-loop oxygen control. The objective of this proof-of-concept study was to determine whether closed-loop control was able to respond to desaturation and subsequent recovery in a controlled laboratory-based environment. METHODS: We conducted a single-blind randomized crossover trial in adults with chronic respiratory disease who had a resting S-pO2 >= 92% and desaturated to < 90% during a 6-min walk test (6MWT). Nasal high-flow was administered during a 6MWT and a subsequent 10-min rest period with either room air, a fixed concentration of 28% oxygen, or oxygen titrated automatically using closed-loop control. RESULTS: The study involved 42 subjects. Closed-loop control maintained S-pO2 within the target range of 92-96% for a mean (SD) duration of 54.4 +/- 30.1% of the 6MWT and 673 +/- 26.8% of the recovery period. The proportion of time spent with an S(pO2 )in the target range during the 6MWT was significantly greater for closed-loop control compared to room air, with a difference of 26.0% (95% CI 17.7-34.2, P < .001); this proportion of time was not significantly different compared to the fixed concentration of 28% oxygen, with a difference of -8.2% (95% CI -16.5 to 0.1, P = .052). The proportion of time spent in the target range during the rest period was significantly greater compared to 28% oxygen, with a difference of 19.3% (95% CI 8.9-29.7, P < .001); this proportion of time was not significantly different compared to room air, with a difference of -9.3% (95% CI -19.7 to 1.0, P = .08). CONCLUSIONS: This study provides proof-of-concept evidence that the novel nasal high-flow device with closed-loop control can respond to changes in S-pO2 outside a target saturation range using a model of exercise-induced desaturation and subsequent recovery.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.2
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据