4.5 Article

Hidden Electrode Failure in a Cochlear Implant User

期刊

LARYNGOSCOPE
卷 131, 期 4, 页码 E1275-E1278

出版社

WILEY
DOI: 10.1002/lary.29023

关键词

Device failure; cochlear implant; ECAP; speech comprehension; impedance

资金

  1. Projekt DEAL

向作者/读者索取更多资源

This study presents a case where abnormal impedance of electrode contacts in a cochlear implant resulted in reduced speech comprehension and distorted sound perception, despite not being detected by conventional tests. Re-implantation significantly improved the patient's subjective sound perception and speech comprehension, indicating the importance of careful monitoring and consideration of technical measurements and hearing performance data in cochlear implant patients.
When the impedance of an electrode contact is highly increased in a cochlear implant, a failure of the appropriate electrode seems obvious. We present a case where impedances of some electrodes were at the lower edge of the normal range and not regarded as suspicious neither by the clinical fitting software nor by in-vivo tests conducted by the implant manufacturer. However, speech comprehension was substantially degraded and sound perception distorted. Also, on the affected electrodes, loudness perception was compromised and responses of the electrically evoked compound action potential were no longer measureable. After re-implantation, the subjective sound percept was clear again and speech comprehension scored much better than before. Later, inspection of the explant revealed shorts on the device and the implant was classified as device failure. Our case shows the importance of collecting longitudinal data of cochlear implant patients, i.e. device related technical measurements and hearing performance data, and the consideration of all these data in cases of patient complaints or suspected implant failures.Laryngoscope, 2020

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据