4.3 Article

Simple phantom fabrication for MRI-based HDR brachytherapy applicator commissioning

期刊

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS
卷 21, 期 11, 页码 283-287

出版社

WILEY
DOI: 10.1002/acm2.13039

关键词

brachytherapy; HDR; image guidance; MRI

向作者/读者索取更多资源

A new high dose rate (HDR) brachytherapy program was initiated in a community hospital setting, with the goal of using magnetic resonance (MR) images with the implant in place during the planning process. Physics acceptance testing and commissioning was completed for key program components, including multiple applicators. To image new applicators for MRI-based planning prior to use with patients, agar gel doped with copper sulfate was created using simple, MR-safe household materials as a practical and inexpensive alternative to custom-machined precision phantoms. Applicators in-phantom were scanned in a 1.5 T MRI scanner using the same sequences developed for the brachytherapy program, then rigidly registered to high-resolution computed tomography (CT) images to assess distortion, artifact, and geometric displacement. To date, Varian tandem and ring sets, segmented cylinders, cervical probes, endometrial applicators; and third-party plastic needles, tandems, and vaginal guides have been imaged in phantom and are available for use clinically.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。