期刊
COMPUTERS IN BIOLOGY AND MEDICINE
卷 69, 期 -, 页码 29-36出版社
PERGAMON-ELSEVIER SCIENCE LTD
DOI: 10.1016/j.compbiomed.2015.12.001
关键词
Aortic root; Finite element analysis; Transcatheter heart valves; Calcification
Background: There are two main methods used for transcatheter aortic valve replacement (TAVR) FEA modeling for medical devices development: patient specific and general approaches. Advantages and disadvantages of both approaches have never been compared in a single study. Method: Here we propose a bioinformatic algorithm to evaluate the accuracy of patient specific and generalized FEA approaches with regards to proximity configuration of the implanted stent reconstructed by computed tomography. In addition, in this study we also assessed the impact of the level of detail on FEA accuracy in the patient specific approach. Results: Our results demonstrate that in certain cases, the generalized approach can ensure the same accuracy as the patient specific approach. Therefore, considering high cost effectiveness of the generalized approach, we identify it as more feasible in the context of TAVR. Furthermore, we suggest that high level of detail can improve the reproducibility of modeling results in the patient specific approach. Conclusions: Our findings may help medical engineers to better understand the peculiarities of both approaches and therefore make the right decision when choosing a particular approach for computer modeling. Future studies are required to validate our observations. (C) 2015 Elsevier Ltd. All rights reserved.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐