CLINICAL CHEMISTRY AND LABORATORY MEDICINE
卷 59, 期 5, 页码 913-920出版社
WALTER DE GRUYTER GMBH
DOI: 10.1515/cclm-2020-0848
关键词
CascadionAnalyzer; cyclosporine A; everolimus; sirolimus; tacrolimus
The study evaluated the analytical performance of the LC-MS/MS panel in comparison to conventional LC-MS/MS and immunoassays, showing excellent performance and the potential for fully automated analysis.
Objectives: Due to its high specificity, liquid chromatographytandem mass spectrometry (LC-MS/MS) is considered the gold standard in diagnostic areas such as therapeutic monitoring of immunosuppressive drugs (ISDs). However, many laboratories still rely on immunoassays for ISD quantification in a tradeoff between analytical performance and the advantages of fully automated analyzers-shorter turnaround times, greater ease of use, and 24/7 availability. Methods: The LC-MS/MS-based Thermo Scientific (TM) Cascadion (TM) SM Immunosuppressant Panel was evaluated for >6 months in the routine laboratory of a university hospital. We assessed the analytical performance of the panel and compared it to conventional LC-MS/MS as well as to immunoassays (cyclosporine A, sirolimus, tacrolimus (Siemens) and everolimus (Thermo Fisher)). In addition, both ISD panel and Cascadion analyzer were scrutinized with regards to, e.g., turnaround time, usability, and robustness. Results: All ISDs showed high linearity and precision (CV=6%) and a good correlation with conventional LC-MS/MS. The mean deviation to the immunoassays was 17-19% and negative for all ISDs except everolimus with a positive 19% bias. No weak points were revealed when challenging assay and system with, e.g., high haematocrit, sedimented whole blood or priority samples. The Cascadion integrated well into our 24/7 routine and could easily be operated simultaneously with several other analyzers by technical staff without LC-MS experience. Conclusions: The ISD panel showed excellent analytical performance and demonstrated that a fully automated LC-MS-based analysis starting from primary samples is feasible, suggesting that LC-MS could become an integral part of 24/7 diagnostics in the near future.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐
暂无数据