4.0 Article

Development of NMIJ CRM 5207-a tungsten dot-array for the image sharpness evaluation in scanning electron microscopy - structure evaluation and determination of dot-pitch

期刊

MICROSCOPY
卷 69, 期 6, 页码 360-370

出版社

OXFORD UNIV PRESS
DOI: 10.1093/jmicro/dfaa030

关键词

SI-traceability; reference material; scanning electron microscopy; image sharpness; resolution; magnification calibration

向作者/读者索取更多资源

We have developed a new certified reference material (CRM) for image sharpness evaluation and magnification calibration for scanning electron microscopy (SEM). Designed to be suitable for the image sharpness evaluation by the derivative method, the CRM has nanoscale tungsten dot-array structure fabricated on silicon substrate, which gives steep contrast transition from the dot to the substrate in SEM image. The pitch of the dot-array was SI-traceably measured as a specified value with relative expanded uncertainty (k = 2) of similar to 1.3%, which can be utilized for the magnification calibration of SEM. Since specimens, as one of the image formation parameters, easily affect image sharpness value, our CRM, as a 'pinned specimen', will play an important role to achieve robust and stable image sharpness measurement system.