4.7 Article

Vascular Stents Coated with Electrospun Drug-Eluting Material: Functioning in Rabbit Iliac Artery

期刊

POLYMERS
卷 12, 期 8, 页码 -

出版社

MDPI
DOI: 10.3390/polym12081741

关键词

paclitaxel; drug-eluting stents; 3D matrix; polycaprolactone; HSA; electrospinning

资金

  1. Russian Scientific Foundation [18-15-00080]
  2. Russian State-funded budget project of ICBFM SB RAS [AAAA-A17-117020210026-2]
  3. Russian Science Foundation [18-15-00080] Funding Source: Russian Science Foundation

向作者/读者索取更多资源

A stenting procedure aimed at blood flow restoration in stenosed arteries significantly improves the efficiency of vascular surgery. However, the current challenge is to prevent neointimal growth, which reduces the vessel lumen, in the stented segments in the long run. We tested in vivo drug-eluting coating applied by electrospinning to metal vascular stents to inhibit the overgrowth of neointimal cells via both the drug release and mechanical support of the vascular wall. The blend of polycaprolactone with human serum albumin and paclitaxel was used for stent coating by electrospinning. The drug-eluting stents (DESs) were placed using a balloon catheter to the rabbit common iliac artery for 1, 3, and 6 months. The blood flow rate was ultrasonically determined in vivo. After explantation, the stented arterial segment was visually and histologically examined. Any undesirable biological responses (rejection or hemodynamically significant stenosis) were unobservable in the experimental groups. DESs were less traumatic and induced weaker neointimal growth; over six months, the blood flow increased by 37% versus bare-metal stents, where it increased by at least double the rate. Thus, electrospun-coated DESs demonstrate considerable advantages over the bare-metal variants.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据