4.5 Article

Simultaneous Monitoring and Comparison of Multiple Product Quality Attributes for Cell Culture Processes at Different Scales Using a LC/MS/MS Based Multi-Attribute Method

期刊

JOURNAL OF PHARMACEUTICAL SCIENCES
卷 109, 期 11, 页码 3319-3329

出版社

ELSEVIER SCIENCE INC
DOI: 10.1016/j.xphs.2020.07.029

关键词

QbD; CQAs trending; Cell culture process; MAM; LC-MS

向作者/读者索取更多资源

In response to the FDA's call for applying Quality by Design (QbD) to the manufacturing process, the biopharmaceutical industry has invested extensively into the monitoring and controlling of product quality attributes for bioprocesses. To assure the safety and efficacy of the drug product, defining critical quality attributes (CQA) and understanding their correlation with critical process parameters (CPP) becomes vitally important. In this work, a liquid chromatography-mass spectrometry based multi-attribute method (MAM) has been applied to the monitoring and trending of multiple CQAs of a monoclonal antibody product. To the best of our knowledge, this is the first demonstration of applying MAM to both a 3-liter development mini-bioreactor (3 L bioreactor) and a 2000-liter GMP single use bioreactor (2000L SUB). MAM was proven not only to be a great analytical tool for monitoring product quality attributes throughout the time course of the cell culture process, it could also provide critical product quality information in order to understand any potential process performance differences during scale-up and/or technology transfer. The successful monitoring and trending of the multiple CQAs throughout the 17-day cell culture process lays a solid foundation for possible real time in-process control and release of biotherapeutics using MAM in the future. (C) 2020 American Pharmacists Association (R). Published by Elsevier Inc. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据