4.2 Article

Fabrication of Carbon Paste Electrodes Modified with Multi-walled Carbon Nanotubes for the Potentiometric Determination of Chromium(III)

期刊

JOURNAL OF ANALYTICAL CHEMISTRY
卷 75, 期 7, 页码 951-957

出版社

PLEIADES PUBLISHING INC
DOI: 10.1134/S1061934820070072

关键词

ion selective electrode; multi-walled carbon nanotubes; D-optimal optimization; potentiometry

向作者/读者索取更多资源

In this research, different electrodes were made by combining certain percentages of carbon paste for a range of cations. Then, ligand selectivity was calculated for each cation using a potentiometric method to identify the most sensitive cation. A modified carbon paste electrode was applied to the quantification of chromium cation based on carbon nanotubes. In addition, paraffin, graphite, and 1,13-bis(8-quinolyl)-l,4,7,10,13-pentaoxatridecane were utilized in the manufacture of the electrodes. To get an optimum composition, D-optimal test design method was employed to increase electrode sensitivity and placement accuracy. Upon combining optimal percentages, this electrode showed the Nernstian response within the concentration range of 1.0 x 10(-10)-1.0 x 10(-2)M, slope of 19.63 mV/decade with a detection limit of 3 x 10(-11)M. The potential was found to remain constant within the pH range of 5.0-9.0, while a gradual deviation was observed outside this range.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.2
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据