Repeatability of whole-cornea measurements using a new swept-source optical coherence tomographer

期刊

EUROPEAN JOURNAL OF OPHTHALMOLOGY
卷 31, 期 4, 页码 1709-1719

出版社

SAGE PUBLICATIONS LTD
DOI: 10.1177/1120672120944022

关键词

Repeatability; optical coherence tomography; swept source Fourier domain; corneal topography; ANTERION

向作者/读者索取更多资源

The study evaluated the repeatability of various corneal parameters provided by a high-resolution SS-OCT. The results showed good repeatability with low Sw and CoV values for most parameters, except for astigmatism. The ICC values were greater than 0.98, confirming the reliability of the measurements.
Purpose: The aim of this study was to evaluate the repeatability of several corneal parameters provided by a high-resolution swept-source optical coherence tomographer (SS-OCT). Methods: One eye from each of 74 patients was measured five times consequently using the ANTERION SS-OCT. The following corneal parameters were analyzed: average, steep and flat keratometry (K), astigmatism for anterior, posterior and total at 3-mm, average K and astigmatism at 6-mm, anterior and posterior eccentricity, higher-order aberrations (HOA) and spherical aberration (Z40), and anterior and posterior best-fit sphere at 8-mm. The intrasubject standard deviation (S-w), coefficient of variation (CoV), coefficient of repeatability (CoR) and intraclass correlation coefficient (ICC) were calculated for each parameter to assess the repeatability. Results: We have not found statistically significant differences between repeated measurements (p> 0.05). Repeatability was good considering the different metrics used. S(w)values were <0.09, varying from 0.035 (posterior average K at 6-mm) to 0.0878 (anterior flat K at 3-mm). CoV values were also low and similar among the different parameters (from 0.08% to 0.21%), except for anterior, posterior and total astigmatism (from 2.25% to 8.46%). Both anterior and posterior eccentricity, and corneal aberrations (HOA andZ40) CoV values were also high. The CoR values were low for all parameters showing those related to the posterior cornea the lowest values (about 0.01). ICC values were >0.98. Conclusions: The ANTERION SS-OCT showed good repeatability when reconstructed different parameters for the whole cornea. This device produces measurements with high repeatability that could be useful for clinical research.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.1
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据