Bimetallic organic framework-based aptamer sensors: a new platform for fluorescence detection of chloramphenicol

期刊

ANALYTICAL AND BIOANALYTICAL CHEMISTRY
卷 412, 期 22, 页码 5273-5281

出版社

SPRINGER HEIDELBERG
DOI: 10.1007/s00216-020-02737-y

关键词

Fluorescence; Sensor; Bimetallic organic framework; Aptamer; Chloramphenicol

资金

  1. National Natural Science Foundation of China [21465010]
  2. Open Foundation of Key Laboratory of Biologic Resources Protection and Utilization of Hubei Province [PKLHB1918, PKLHB1730]
  3. Hubei Provincial Natural Science Foundation of China [2018CFB247]
  4. Special Funds for Double First-Class Construction in Hubei Province

向作者/读者索取更多资源

A fluorescence method for the quantitative detection of chloramphenicol (CAP) has been developed using phosphate and fluorescent dye 6-carboxy-x-rhodamine (ROX) double-labeled aptamers of CAP and the bimetallic organic framework nanomaterial Cu/UiO-66. Cu/UiO-66 was prepared by coordinate bonding of metal organic framework (MOF) nanomaterial UiO-66 with copper ions. Cu/UiO-66 contains a large number of metal defect sites, which can be combined with phosphate-modified nucleic acid aptamers through strong coordination between phosphate and zirconium to form fluorescence turn-on sensors. In the absence of CAP, all single-stranded aptamers were adsorbed on the surface of Cu/UiO-66 through pi-pi stacking between single-stranded DNA and Cu/UiO-66, which brings the ROX fluorophores and Cu/UiO-66 into close proximity. The ROX fluorescence of aptamers was then quenched by Cu/UiO-66 through photoinduced electron transfer (PET). In the presence of CAP, however, CAP reacted with nucleic acid aptamers to form a special spatial structure, in which the ROX fluorophores were far away from the MOF surface via a change in the spatial structure of the aptamers, and the fluorescence of ROX was able to be recovered. The quantitative detection of CAP can be achieved by measuring the fluorescence signal of ROX using synchronous scanning fluorescence spectrometry. Under optimum conditions, the fluorescence intensities of ROX exhibit a good linear dependence on the concentration of CAP in the range of 0.2-10 nmol/L, with a detection limit of 0.09 nmol/L. The method has advantages of high sensitivity, good selectivity, and a low limit of detection.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据