期刊
JOURNAL OF SYNCHROTRON RADIATION
卷 27, 期 -, 页码 813-826出版社
INT UNION CRYSTALLOGRAPHY
DOI: 10.1107/S160057752000243X
关键词
X-ray instrumentation; wavelength dispersive spectrometer; X-ray optics; tender X-rays; Johansson crystal analyzers
资金
- French National Research Agency (ANR) [ANR-10-EQPX-27-01]
The design and first results of a large-solid-angle X-ray emission spectrometer that is optimized for energies between 1.5 keV and 5.5 keV are presented. The spectrometer is based on an array of 11 cylindrically bent Johansson crystal analyzers arranged in a non-dispersive Rowland circle geometry. The smallest achievable energy bandwidth is smaller than the core hole lifetime broadening of the absorption edges in this energy range. Energy scanning is achieved using an innovative design, maintaining the Rowland circle conditions for all crystals with only four motor motions. The entire spectrometer is encased in a high-vacuum chamber that allocates a liquid helium cryostat and provides sufficient space for in situ cells and operando catalysis reactors.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐