Proposal of a low-dose, long-pitch, dual-source chest CT protocol on third-generation dual-source CT using a tin filter for spectral shaping at 100 kVp for CoronaVirus Disease 2019 (COVID-19) patients: a feasibility study

期刊

RADIOLOGIA MEDICA
卷 125, 期 4, 页码 365-373

出版社

SPRINGER-VERLAG ITALIA SRL
DOI: 10.1007/s11547-020-01179-x

关键词

COVID-19; 2019-nCOV; Dual-source CT; Spectral shaping; Low-dose CT; Chest radiology

向作者/读者索取更多资源

Aim To subjectively and objectively evaluate the feasibility and diagnostic reliability of a low-dose, long-pitch dual-source chest CT protocol on third-generation dual-source CT (DSCT) with spectral shaping at 100Sn kVp for COVID-19 patients. Materials and methods Patients with COVID-19 and positive swab-test undergoing to a chest CT on third-generation DSCT were included. The imaging protocol included a dual-energy acquisition (HD-DECT, 90/150Sn kVp) and fast, low-dose, long-pitch CT, dual-source scan at 100Sn kVp (LDCT). Subjective (Likert Scales) and objective (signal-to-noise and contrast-to-noise ratios, SNR and CNR) analyses were performed; radiation dose and acquisition times were recorded. Nonparametric tests were used. Results The median radiation dose was lower for LDCT than HD-DECT (Effective dose, ED: 0.28 mSv vs. 3.28 mSv, p = 0.016). LDCT had median acquisition time of 0.62 s (vs 2.02 s, p = 0.016). SNR and CNR were significantly different in several thoracic structures between HD-DECT and LDCT, with exception of lung parenchyma. Qualitative analysis demonstrated significant reduction in motion artifacts (p = 0.031) with comparable diagnostic reliability between HD-DECT and LDCT. Conclusions Ultra-low-dose, dual-source, fast CT protocol provides highly diagnostic images for COVID-19 with potential for reduction in dose and motion artifacts.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据