4.6 Article

A comparison of automated and manual co-registration for magnetoencephalography

期刊

PLOS ONE
卷 15, 期 4, 页码 -

出版社

PUBLIC LIBRARY SCIENCE
DOI: 10.1371/journal.pone.0232100

关键词

-

资金

  1. National Institute on Alcohol Abuse and Alcoholism [R01AA023665, K01AA021431]

向作者/读者索取更多资源

Magnetoencephalography (MEG) is a neuroimaging technique that accurately captures the rapid (sub-millisecond) activity of neuronal populations. Interpretation of functional data from MEG relies upon registration to the participant's anatomical MRI. The key remaining step is to transform the participant's MRI into the MEG head coordinate space. Although both automated and manual approaches to co-registration are available, the relative accuracy of two approaches has not been systematically evaluated. The goal of the present study was to compare the accuracy of manual and automated co-registration. Resting MEG and T1-weighted MRI data were collected from 90 participants. Automated and manual co-registration were performed on the same subjects, and the inter-method reliability of the two methods assessed using the intra-class correlation. Median co-registration error for both methods was within acceptable limits. Inter-method reliability was in the good range for co-registration error, and the good to excellent range for translation and rotation. These results suggest that the output of the automated co-registration procedure is comparable to that achieved using manual co-registration.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据