4.5 Article

Surgeon-defined assessment is a poor predictor of knee balance in total knee arthroplasty: a prospective, multicenter study

期刊

出版社

SPRINGER
DOI: 10.1007/s00167-020-05925-6

关键词

Total knee arthroplasty; Balance; Assessment; Sensors; Soft tissue; Pressure

资金

  1. OrthoSensor (Dania Beach, FL, USA)
  2. Ramsay Hospital Research Foundation (Sydney, Australia)

向作者/读者索取更多资源

This study found that surgeon-defined assessment of soft tissue balance is a poor predictor of balance in total knee arthroplasty, especially in unbalanced cases. Surgeon experience did not affect test accuracy, and there was considerable variability between different surgeons. The accuracy of SDA in TKA is questioned based on these findings.
Background The accuracy of surgeon-defined assessment (SDA) of soft tissue balance in total knee arthroplasty (TKA) is poorly understood despite balance being considered a significant determinant of surgical success. The study's hypothesis was that intra-operative SDA is a poor predictor of coronal balance in TKA. Methods A prospective, multicenter study assessing accuracy of SDA of balance was conducted in 250 patients (285 TKAs). Eight surgeons and thirteen trainees participated, and all were blinded to sensor measurements. The primary outcome was test accuracy of SDA measured at 10 degrees, 45 degrees and 90 degrees compared to sensor measures as the gold standard test. Cohen's kappa coefficient was calculated to determine chance-corrected agreement. Secondary outcomes include the relationship of SDA to level of surgical experience, analysis of between-surgeon differences, and the influence of patient and operative factors on SDA accuracy. Results Average accuracy of SDA was 58.3%, 61.2% and 66.5% at 10 degrees, 45 degrees and 90 degrees respectively. Cohen's kappa coefficient was 0.18 at all angles and rated as slight agreement. SDA sensitivities to correctly identify a balanced knee (76.2% at 10 degrees; 82.6% at 45 degrees; 83.2% at 90 degrees) were approximately twice specificities to correctly identify an unbalanced knee (42.6% at 10 degrees; 34.1% at 45 degrees; 41.4% at 90 degrees). Surgical experience (surgeon versus trainee) had no effect on capacity to determine balance. Considerable between-surgeon variability was found (33-65% at 10 degrees, 41-73% at 45 degrees, 55-89% at 90 degrees). Conclusion SDA was a poor predictor of balance, particularly when assessing the unbalanced TKA. Surgeon experience had no effect on test accuracy and considerable between-surgeon variability was recorded. These findings question the accuracy of SDA in TKA. Trial Registration Number: ACTRN# 12618000817246.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据