Analytical performances of a chemiluminescence immunoassay for SARS-CoV-2 IgM/IgG and antibody kinetics

期刊

CLINICAL CHEMISTRY AND LABORATORY MEDICINE
卷 58, 期 7, 页码 1081-1088

出版社

WALTER DE GRUYTER GMBH
DOI: 10.1515/cclm-2020-0443

关键词

analytical performances; antibody kinetics; COVID-19; COVID IgG and IgM; rRT-PCR; SARS-CoV-2

向作者/读者索取更多资源

Background: Coronavirus disease 2019, abbreviated to COVID-19, represents an emerging health threat worldwide as, after initial reports in China, it has continued to spread rapidly. The clinical spectrum of the disease varies from mild to severe acute respiratory distress syndrome (ARDS). Moreover, many patients can be asymptomatic, thus increasing the uncertainty of the diagnostic work-up. Laboratory tests play a pivotal role in the diagnosis and management of COVID-19, the current gold standard being real-time reverse transcription polymerase chain reaction (rRT-PCR) on respiratory tract specimens. However, the diagnostic accuracy of rRT-PCR depends on many preanalytical and analytical variables. The measurement of specific COVID-19 antibodies (both IgG and IgM) should serve as an additional, non-invasive tool for disease detection and management. Methods: The imprecision of the MAGLUMI (TM) 2000 Plus 2019-nCov IgM and IgG assays (Snibe, Shenzhen, China) was assessed by adopting the Clinical and Laboratory Standards Institute (CLSI) EP15-A3 protocol. Linearity of dilution and recovery was evaluated by means of mixes of high-level pools and low-level pools of serum samples. Immunoglobulin time kinetics were evaluated using a series of serum samples, repeatedly collected from COVID-19-positive patients at different times, from <5 days up to 26-30 days. Results: Findings at the analytical validation of the assay carried out according to the CLSI EP15-A3 guideline demonstrated that imprecision and repeatability were acceptable (repeatability was <4% and <6% for IgM and IgG, respectively, whilst intermediate imprecision was <6%). In addition, results of dilution and recovery studies were satisfactory. The kinetics of COVID-19 antibodies confirmed previously reported findings, showing a rapid increase of both IgM and IgG after 6-7 days from the symptom onset. IgG had 100% sensitivity on day 12, whilst 88% was the higher positive rate achieved for IgM after the same time interval. Conclusions: The findings of this study demonstrate the validity of the MAGLUMI 2000 Plus CLIA assay for the measurement of specific IgM and IgG in sera of COVID-19 patients, and for obtaining valuable data on the kinetics of both (IgM and IgG) COVID-19 antibodies. These data represent a pre-requisite for the appropriate utilization of specific antibodies for the diagnosis and management of COVID-19 patients.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据