期刊
CLINICAL CHEMISTRY AND LABORATORY MEDICINE
卷 58, 期 8, 页码 1191-1199出版社
WALTER DE GRUYTER GMBH
DOI: 10.1515/cclm-2020-0573
关键词
COVID-19; D-dimer; thrombosis; units
Coronavirus disease 2019 (COVID-19) represents a new pandemic caused by severe acute respiratory syndrome virus coronavirus 2 (SARS-CoV-2). A previous pooled analysis clearly identified elevated D-dimer levels as being associated with severity of COVID-19. Since then, several other studies have provided clearer support for this initial evidence. However, potentially under-recognized by those reporting on D-dimer is the considerable variation in reporting units for D-dimer, and thus also the potential for misreporting of D-dimer data based on poor or incomplete reporting. A PubMed search was used to identify recent papers reporting on D-dimers in COVID-19-based studies. We report that: (1) most publications did not identify either the manufacturer or D-dimer product used; (2) most did not identify whether D-dimer values were reported as D-dimer units (DDU) or fibrinogen equivalent units (FEU) (similar to 2 x differences); (3) nearly half did not identify normal cut-off values; (4) some did not report numerical findings or units for D-dimer; (5) where reported, most identified units as either mg/L or mu g/mL; (6) we identified at least four errors in reporting from 21 papers. It may not be possible to truly standardize D-dimer assays, but it should be feasible to harmonize D-dimer assays to a single unit of measurement.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐