Determining patient abdomen thickness from a single digital radiograph with a computational model: clinical results from a proof of concept study

期刊

BRITISH JOURNAL OF RADIOLOGY
卷 93, 期 1111, 页码 -

出版社

BRITISH INST RADIOLOGY
DOI: 10.1259/bjr.20200010

关键词

-

向作者/读者索取更多资源

Objective: A computational model has been created to estimate the abdominal thickness of a patient following an X-ray examination; its intended application is assisting with patient dose audit of paediatric X-ray examinations. This work evaluates the accuracy of the computational model in a clinical setting for adult patients undergoing anteroposterior (AP) abdomen X-ray examinations. Methods: The model estimates patient thickness using the radiographic image, the exposure factors with which the image was acquired, a priori knowledge of the characteristics of the X-ray unit and detector and the results of extensive Monte Carlo simulation of patient examinations. For 20 patients undergoing AP abdominal X-ray examinations, the model was used to estimate the patient thickness; these estimates were compared against a direct measurement made at the time of the examination. Results: Estimates of patient thickness made using the model were on average within +/- 5.8% of the measured thickness. Conclusion: The model can be used to accurately estimate the thickness of a patient undergoing an AP abdominal X-ray examination where the patient's size falls within the range of the size of patients used to create the computational model. Advances in knowledge: This work demonstrates that it is possible to accurately estimate the AP abdominal thickness of an adult patient using the digital X-ray image and a computational model.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据