The study assessed the repeatability and reliability of tomographic parameters for characterizing keratoconus measured with a Pentacam HR. Results showed excellent ICCs for all parameters, with repeatability limits varying for different parameters, supporting the usefulness of Pentacam HR for diagnosing keratoconus. New criteria should be established for monitoring progression based on real measurements from this system.
Purpose To assess the repeatability and reliability of the most important tomographic parameters for characterising keratoconus measured with a Pentacam HR (high resolution). Methods Overall, 230 eyes in 158 patients with keratoconus were analysed. We performed five consecutive corneal tomography examinations for each eye with a Pentacam HR in patients with keratoconus. Study eyes were classified into three groups depending on the maximum posterior elevation (max_BFS_post): grade 1 for cases of keratoconus with a max_BFS_post of 40 mu m; grade 2 for those with a max_BFS_post of between 41 and 75 mu m and grade 3 for those with a max_BFS_post of over 75 mu m. We calculated the intraclass correlation coefficients (ICCs) and repeatability limits of parameters from tomography and aberrometry. Results All the parameters were found to have excellent ICCs (0.9). The repeatability limits for the key parameters were higher than 0.5D for the power parameters, 20 degrees for the axis of corneal astigmatism and 10 mu m for the thinnest corneal thickness. Further, we obtained repeatability limits of above 0.1 mu m for the aberrometry values and overall greater than 15 degrees for the coma axis. All the values increase with the severity of keratoconus, except for that of the coma axis which falls with keratoconus grade. Conclusions The reliability indicated by ICCs supports the view that the Pentacam HR is useful for the diagnosis of keratoconus. The repeatability limits suggest that new criteria should be established for monitoring progression taking into account the real measurements that can be made using this system.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据