4.0 Article

Validation of the Spacelabs 90227 OnTrak device according to the European and British Hypertension Societies as well as the American protocols

期刊

BLOOD PRESSURE MONITORING
卷 25, 期 2, 页码 110-114

出版社

LIPPINCOTT WILLIAMS & WILKINS
DOI: 10.1097/MBP.0000000000000424

关键词

American National Standards Institute; Association for the Advancement of Medical Instrumentation; International Standards Organization protocol; automated blood pressure measurement; British Hypertension Society; device validation; European Hypertension Society

资金

  1. Spacelabs Healthcare

向作者/读者索取更多资源

Objective To assess the static device accuracy of the Spacelabs 90227 OnTrak according to the three most recognized validation protocols, in both adults and children. Methods Subjects were recruited and data gathered and analysed according to the European Society of Hypertension (ESH-IP2), the British Hypertension Society (BHS) and the American National Standards Institute/Association for the Advancement of Medical Instrumentation/International Standards Organization (ANSI/AAMI/ISO) protocols. Results The device achieved a mean difference +/- SD between observers and device for SBP/DBP was 0.9 +/- 7/-1.4 +/- 6 mmHg (ESH-IP2) and 1.7 +/- 8/-1.5/6 mmHg (BHS) Grade A/A. For the ANSI/AAMI/ISO protocol, the average device-observer difference for SBP/DBP was 0.4 +/- 7/-1.5 +/- 7 mmHg fulfilling the two criteria of the protocol. Conclusion The Spacelabs 90227 OnTrak device achieved the criteria of all three recognised validation protocols and can therefore be recommended for clinical use as a static device.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.0
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据