4.3 Article

Effects of Mechanical Vibration during Drying of Solution Sample in Enhancing the Sensitivity of Total Reflection X-ray Fluorescence Analysis

期刊

ANALYTICAL SCIENCES
卷 36, 期 4, 页码 465-469

出版社

SPRINGERNATURE
DOI: 10.2116/analsci.19P372

关键词

Coffee ring; dry residue; total reflection X-ray fluorescence; vanadium; vibration

向作者/读者索取更多资源

This study describes a sample preparation method for trace element analysis using total reflection X-ray fluorescence analysis. The procedure of this method is as follows: 1) a hydrophobic-coated sample holder whose hydrophobic film at around the center of the surface is partially dissolved by acetone is prepared; 2) a droplet of a sample solution is dropped on the sample holder; and 3) the sample droplet is heated and dried while vibrating it. This sample preparation method is effective for reducing the size of the dry residue of the sample solution and suppressing the coffee-ring formation. Furthermore, this method leads to an enhancement in the intensity of an X-ray fluorescence line from an analyte element in the sample residue. In this study, this sample preparation method led to an improvement in the detection limit for vanadium in commercially available bottled drinking water.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.3
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据