Determination of 117 endocrine disruptors (EDCs) in water using SBSE TD-GC-MS/MS under the European Water Framework Directive

期刊

ANALYTICAL AND BIOANALYTICAL CHEMISTRY
卷 412, 期 21, 页码 5169-5178

出版社

SPRINGER HEIDELBERG
DOI: 10.1007/s00216-020-02553-4

关键词

SBSE; Endocrine disruptors; Dynamic multiple reaction monitoring (dMRM); EWFD; Water analysis

向作者/读者索取更多资源

The aim of this work is the highly sensitive determination of 117 contaminants in surface waters at very low concentrations. Gas chromatography-amenable priority compounds from the European Water Framework Directive (EWFD, 45 priority substances) and substances from the Turkish Regulation on the Management of Surface Water Quality (TRMSWQ, 250 substances) were analyzed. Almost a third to one half of the compounds in the TRMSWQ list were successfully detected in a single gas chromatography-tandem mass spectrometry (GC-MS/MS) injection. Analyzing these pollutants in water samples is difficult with conventional extraction methods. An efficient, easy and fast method for both extraction and analysis is thus important. Here, the stir bar sorptive extraction (SBSE) approach employing Twister(R) was used for analyte enrichment from 100 mL water samples. The Twister apparatus was used for thermal desorption. Compounds were separated via gas chromatography (GC) and detected by tandem mass spectrometry (MS/MS) utilizing an Agilent 7000D MS instrument. The analysis method was comprehensively validated, and complied with the requirements of the EWFD and Turkish regulation for surface water quality for inland surface water. The method includes various endocrine disruptor compounds listed in the EWFD, such as polychlorinated biphenyls, polycyclic aromatics hydrocarbons, polybrominated diphenyl ethers, phenols, phthalates and pesticides. The method is also applicable for the analysis of similar contaminants that are not on this list. The analyzed pollutants have varying polarities based on octanol-water coefficients and are extracted by SBSE. This method may be preferred for the analysis of pollutants with an octanol-water partition coefficient (log Ko/w) higher than 2. The SBSE method, which is easy, lab-friendly and quick, taking a total of only 2 h for parameter analysis, was optimized to reduce time and chemical usage. The entire extraction and analysis was completed in virtually 3 h. The limits of quantification (LOQs) ranged from 0.12 to 50 ng/L and complied with the requirements for analytical methods to be used in the analysis of the compounds included in the directives. Several quality parameters including linearity, trueness and precision were studied, with good results, and uncertainty was also estimated. Precision (in terms of relative standard deviation [RSD]) was lower than 40%, recoveries ranged from 60 to 120%, and determination coefficients were higher than 0.990 for all analytes. Graphical abstract

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据