Fluid-Structure Coupling Model and Experimental Validation of Interaction Between Pneumatic Soft Actuator and Lower Limb

期刊

SOFT ROBOTICS
卷 7, 期 5, 页码 627-638

出版社

MARY ANN LIEBERT, INC
DOI: 10.1089/soro.2019.0035

关键词

pneumatic soft actuator; compression therapy; fluid-structure coupling model; experimental validation; lower limb

类别

资金

  1. Innovation and Technology Fund of Hong Kong SAR Government [ITS/031/17]
  2. Central Research Grant of the Hong Kong Polytechnic University [1-ZVLQ]
  3. University Grants Committee [1-ZE7K]
  4. Natural Science Foundation of Jiangsu Province [BK20180933]
  5. Natural Science Foundation of Jiangsu Higher Institutions [19KJB460028]
  6. QingLan Project of Yangzhou University

向作者/读者索取更多资源

Pneumatic soft actuators (PSAs) are components that produce predesigned motion or force in different end-use devices. PSAs are lightweight, flexible, and compatible in human-machine interaction. The use of PSAs in compression therapy has proven promising in proactive pressure delivery with a wide range of dosages for treatment of chronic venous insufficiency and lymphedema. However, effective design and control of PSAs for dynamic pressure delivery have not been fully elaborated. The purpose of this study is to explore interactive working mechanisms between a PSA and lower limbs through establishing fluid-structure coupling models, an intermittent pneumatic compression (IPC) testing system, and conducting experimental validation. The developed IPC testing system consisted of a PSA unit (multichambered bladders laminated with an external textile shell), a pneumatic controller, and various real-time pressure monitoring sensors and accessory elements. The established coupling model characterized the dynamic response process with varying design parameters of the PSA unit, and demonstrated that the design of initial thickness, stiffness, and air mass flow of the PSA, as well as stiffness of limb tissues of the users, influenced PSA-lower limb interactions and resultant pressure dosages. The simulated results presented a favorable agreement with the experimental data collected by the IPC testing system. This study enhanced understanding of PSA-lower limb interactive working mechanisms and provided an evidence-based technical guidance for functional design of PSA. These results contribute to improving the efficacy of dynamic compression therapy for promotion of venous hemodynamics and user compliance in practice.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据