Ionic liquid-based reference electrodes for miniaturized ion sensors: What can go wrong?

期刊

SENSORS AND ACTUATORS B-CHEMICAL
卷 301, 期 -, 页码 -

出版社

ELSEVIER SCIENCE SA
DOI: 10.1016/j.snb.2019.127112

关键词

Reference electrode; Ionic liquid Salt Bridge; Potentiometry; Ion-selective electrode; Liquid junction potential; Ion sensing

资金

  1. Wyss Institute for Biologically Inspired Engineering
  2. Harvard REU program under NSF [DMR-1420570]
  3. Trinity Hall, University of Cambridge
  4. NSF [CHE-1710024]

向作者/读者索取更多资源

Ionic liquid-based reference electrodes, especially those containing 1-octyl-3-methylimidazolium bis(tri-fluoromethylsulfonyl)imide (MeOctIm TFSI), are increasingly used in potentiometric measurements. They replace the conventional salt bridges and establish the reference potential through local partitioning of the ions of the ionic liquid between the reference membrane and the sample solution. Unless the electrochemical cell and the measurement protocol are designed appropriately, these ions can interfere with the response of ion-selective electrodes (ISEs) with polymeric membranes. This work characterizes the effect of MeOctIm TFSI on the response of K+, Na+, and Cl- ISEs with polymeric membranes. The leaching of MeOctIm TFSI from the reference membrane to the sample solution was monitored with UV-vis spectroscopy. The concentration of MeOctIm TFSI in the aqueous phase increased gradually and plateaued at approximately 1.0 mM. Concentrations as low as 100 mu M of MeOctIm TFSI caused large changes to the emf (50-150 mV) of K+, Na+, and Cl- ISEs. The presence of 10 mu M of TESI- was enough to cause Donnan failure of K+ ISEs with valinomycin as ionophore (that is, co-extraction of and K+ TESI- into the sensing membrane). Use of less lipophilic anions such as tetrafluoroborate (BF4-) or triflate (OTf-) postponed the onset of Donnan failure of cation-selective ISEs to higher concentrations of the anion, but decreased the stability in the reference potential and lifetime of the reference electrode. These results imply that although MeOctIm TFSI-based reference electrodes provide sample-independent and stable electrical potentials, they should be used with caution for measurements with polymeric-membrane ISEs, due to strong interference of both MeOctIm(+) and TESI- with measured values of emf.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据