4.4 Article

High-performance thin-layer chromatography with atmospheric solids analysis probe mass spectrometry for analysis of gasoline polymeric additives

期刊

出版社

WILEY
DOI: 10.1002/rcm.8755

关键词

-

资金

  1. Labex SynOrg [ANR-11-LABX-0029]
  2. European Regional Development Fund [HN0001343]

向作者/读者索取更多资源

Rationale The offline coupling of high-performance thin-layer chromatography (HPTLC) with atmospheric solids analysis probe mass spectrometry (ASAP-MS) was evaluated for the characterization of polymeric additives in gasoline. Methods A protocol was developed to optimize the ion signal. A glass capillary was moistened with deionized water, and then dipped into silica gel scratched from an HPTLC plate. The capillary tube was fixed to the ASAP holder and introduced into the ionization source for analysis by MS. Silica gel, reversed-phase C18 and cellulose stationary phases were evaluated. Results The effect of the stationary phase and the nature of analyte were evaluated using polypropylene glycol and polyisobutylene succinimide polyamine as analyte molecules. The optimal ionization conditions are significantly different between ASAP and HPTLC/ASAP-MS analyses. In particular, a higher desorption gas temperature was required to produce ions from the silica gel HPTLC plate. The presence of the stationary phase reduces the internal energy of the ions and limits the fragmentation. Conclusions HPTLC/ASAP-MS is a very fast and efficient technique for the analysis of polymers in formulated fuels. Good ionization efficiency was obtained with all investigated stationary phases.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据