Pre-clinical evaluation of dual-layer spectral computed tomography-based stopping power prediction for particle therapy planning at the Heidelberg Ion Beam Therapy Center

期刊

PHYSICS IN MEDICINE AND BIOLOGY
卷 65, 期 9, 页码 -

出版社

IOP PUBLISHING LTD
DOI: 10.1088/1361-6560/ab735e

关键词

dual-layer spectral CT; stopping power; particle therapy; range uncertainties

向作者/读者索取更多资源

Accurate beam range prediction during clinical treatment planning is key to improve targeted dose delivery in proton and heavy ion therapy. A substantial source of beam range uncertainty is the prediction of ion stopping power ratio (SPR) relative to water from an empirical calibration based on conventional x-ray computed tomography (CT) used in clinical practice today. The aim of this study was to investigate the potential of a novel spectral CT imaging technique based on a dual-layer detector-based approach to improve the SPR prediction for particle therapy treatment planning, an improvement that would minimize the beam range uncertainty and allow for reduced safety margins in the patient. Using calibrated and validated maps of electron density and effective atomic number from spectral CT data, predicted SPR values in tissue substitutes were within a mean accuracy of 0.6% compared to measured SPR and showed substantially better agreement with measured data compared to standard CT-number-to-SPR calibration. The accuracy of SPR was not affected by CT acquisition settings, reconstruction parameters, phantom size, and type. Additionally, various spectral CT conversion algorithms were compared to determine the most accurate prediction method. Dosimetric validation of the developed method using a half-head anthropomorphic phantom in a routine-like setting indicated that SPR prediction with dual-layer spectral CT outperforms the clinical single-energy CT standard with a range prediction improvement of 1 mm. This study demonstrated in homogeneous and heterogeneous phantoms that spectral CT is feasible for particle therapy planning to improve range estimates for high-precision particle therapy.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据