Recent research has shown promising results for the development of a clinically feasible vestibular implant in the near future. However, correct electrode placement remains a challenge. It was shown that fluoroscopy was able to visualize the semicircular canal ampullae and electrodes, and guide electrode insertion in real time. Ninety-four percent of the 18 electrodes were implanted correctly (<1.5 mm distance to target). The median distances were 0.60 mm, 0.85 mm, and 0.65 mm for the superior, lateral, and posterior semicircular canal, respectively. These findings suggest that fluoroscopy can significantly improve electrode placement during vestibular implantation.