4.8 Article

A comparative study of different protein immobilization methods for the construction of an efficient nano-structured lactate oxidase-SWCNT-biosensor

期刊

BIOSENSORS & BIOELECTRONICS
卷 64, 期 -, 页码 138-146

出版社

ELSEVIER ADVANCED TECHNOLOGY
DOI: 10.1016/j.bios.2014.08.072

关键词

Covalent immobilization; Lactate biosensor; Lactate oxidase; Physical adsorption; Protein immobilization; Protein stability

资金

  1. National Institutes of Health (NIH) [SC1 GM086240]
  2. NASA URC Grant [NNG05GG78H]
  3. Puerto Rico NASA Space Grant Consortium
  4. NASA Jenkins Pre-doctoral Fellowship Project (JPFP)
  5. Office Of The Director
  6. Office of Integrative Activities [1002410] Funding Source: National Science Foundation

向作者/读者索取更多资源

We constructed lactate biosensors by immobilization of lactate oxidase (LOx) onto a single-walled carbon nanotube (SWCNT) electrode. The first step of the sensor construction was the immobilization of oxidized SWCNT onto a platinum electrode modified with 4-aminothiophenol (4-ATP). Two enzyme immobilization methods were used to construct the biosensors, i.e., covalent immobilization using 1-ethyl-3-(3-dimethylaminopropyl) carbodiimide hydrochloride (EDC) and physical adsorption. Atomic force microscopy (AFM) experiments confirmed the immobilization of SWCNT during the biosensor construction and X-ray photoelectron spectroscopy (XPS) experiments confirmed covalent immobilization of LOx onto the SWCNT in the first method. The biosensor based on covalent enzyme immobilization showed a sensitivity of 5.8 mu A/mM, a linearity up to 0.12 mM of L-lactate, and a detection limit of 4.0 mu M. The biosensor based on protein adsorption displayed a sensitivity of 9.4 mu A/mM, retaining linearity up to 0.18 mM of t-lactate with a detection limit of 3.0 11M. The difference in the biosensor response can be attributed to protein conformational or dynamical changes during covalent immobilization. The stability of the biosensors was tested at different temperatures and after different storage periods. The thermostability of the biosensors after incubation at 60 degrees C demonstrated that the biosensor with covalently immobilized LOx retained a higher response compared with the adsorbed protein. Long-term stability experiments show a better residual activity of 40% for the covalently immobilized protein compared to 20% of residual activity for the adsorbed protein after 25 d storage. Covalent protein immobilization was superior compared to adsorption in preserving biosensor functionality over extended time period. (C) 2014 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.8
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据