4.7 Article

Proposition of a periodic verification test for Articulated Arm Coordinate Measuring Machines using a small 3D artefact

期刊

MEASUREMENT
卷 154, 期 -, 页码 -

出版社

ELSEVIER SCI LTD
DOI: 10.1016/j.measurement.2020.107472

关键词

AACMM; Performance verification; ISO 10360-12; Kinematic redundancy

向作者/读者索取更多资源

This paper proposes a periodic verification/acceptance test for Articulated Arm Coordinate Measuring Machine (AACMM). The verification tests recommended by standards dedicated to AACMM are generally time consuming and require bulky artifacts. In fact, the volumetric test of ISO 10360-12 requires measuring five lengths in seven different directions, ensuring that the largest length covers at least 66% of the working range. The proposed test in the paper requires only small artifacts and takes 20-30 min. The test is designed in a way to move all encoders of the AACMM through their entire angular range, with measurements located in a small area of the working range of the AACMM, thanks to the redundancy of the kinematic structure of AACMMs. To validate the proposed test, a comparison with the ISO 10360-12 volumetric test (in terms of encoders' error, mechanical deformation errors and the maximal error captured) is presented. (C) 2020 Elsevier Ltd. All rights reserved.