Comparison of cord blood lactate measurement by gas analyzer and portable electrochemical devices

期刊

JOURNAL OF PERINATAL MEDICINE
卷 48, 期 2, 页码 157-161

出版社

WALTER DE GRUYTER GMBH
DOI: 10.1515/jpm-2019-0357

关键词

acid base status; base excess; cord blood lactate; fetal hypoxia; gas analyzer; labor; partial carbon dioxide pressure; partial oxygen pressure; pH

向作者/读者索取更多资源

Objective: To compare the accuracy of cord blood lactate measurement using gas analyzer and portable devices in order to assess possibility of implementation of these devices in clinical practice. Methods: We performed a prospective observational study using 30 umbilical cord samples which were obtained immediately after birth. Portable electrochemical devices Lactate Scout (Sens Lab GmbH, Leipzig, Germany) and StatStrip Lactate (NOVA Biomedical, Waltham, MA, USA) were used to determine lactate level. A gas analyzer ABL800 FLEX (Radiometer Medical ApS, Bronshoj-Husum, Denmark) was used as a reference. Base excess (BE), pH, partial oxygen (pO(2)) and carbon dioxide (pCO(2)) pressure, hemoglobin (ctHb) and bilirubin (ctBl) levels were measured. Results: The mean umbilical cord blood lactate level determined by the gas analyzer was 5.85 +/- 2.66 mmol/L (ranging from 1.4 mmol/L to 13.4 mmol/L). Lactate level estimated by Lactate Scout was 5.66 +/- 2.65 mmol/L and did not significantly differ from the reference method level (P= 0.2547). The mean lactate level determined by StatStrip Lactate was significantly lower than by the gas analyzer - 4.81 +/- 2.38 mmol/L (P < 0.0001). Umbilical cord blood pH, BE, pO(2) and pCO(2), ctHb and ctBl levels did not affect the accuracy of the lactate measurement in absolute units (mmol/L). Conclusion: Umbilical cord blood lactate level measured by StatStrip Lactate was lower than estimated by the ABL800 FLEX gas analyzer. This shows the necessity to develop decision-making reference points separately for each device. Umbilical cord blood pH, BE, pO(2) and pCO(2), ctHb and ctBl levels did not affect the accuracy of measurements by electrochemical portable devices.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.3
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据