4.3 Review

Post-contrast 3D T1-weighted TSE MR sequences (SPACE, CUBE, VISTA/BRAINVIEW, isoFSE, 3D MVOX): Technical aspects and clinical applications

期刊

JOURNAL OF NEURORADIOLOGY
卷 47, 期 5, 页码 360-370

出版社

MASSON EDITEUR
DOI: 10.1016/j.neurad.2020.01.085

关键词

Gadolinium; Gradient echo imaging; Post-contrast MR-imaging; Variable flip angle 3D TSE sequence

向作者/读者索取更多资源

Post-contrast three-dimensional T1-weighted imaging of the brain is widely used for a broad range of vascular, inflammatory or tumoral diseases. The variable flip angle 3D TSE sequence is now available from several manufacturers (CUBE, General Electric; SPACE, Siemens; VISTA/BRAINVIEW, Philips; isoFSE, Itachi; 3D MVOX, Canon). Compared to gradient-echo (GRE) techniques, 3D TSE offers the advantages of useful image contrasts and reduction of artifacts from static field inhomogeneity. However, the respective role of 3D TSE and GRE MR sequences remains to be elucidated, particularly in the setting of post-contrast imaging. The purpose of this review was (1) to describe the technical aspects of 3D TSE sequences, (2) to illustrate the main clinical applications of the post-contrast 3D T1-w TSE sequence through clinical cases, (3) to discuss the respective role of post-contrast 3D TSE and GRE imaging in the field of neuroimaging. (C) 2020 Elsevier Masson SAS. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.3
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据