4.3 Article

Evaluation of the IAEA-TRS 483 protocol for the dosimetry of small fields (square and stereotactic cones) using multiple detectors

期刊

出版社

WILEY
DOI: 10.1002/acm2.12792

关键词

output factors; small field dosimetry; stereotactic cones; TRS-483

向作者/读者索取更多资源

The IAEA TRS 483 protocol(1) for the dosimetry of small static fields in radiotherapy was used to calculate output factors for the Elekta Synergy linac at the Australian Radiation Protection and Nuclear Safety Agency (ARPANSA). Small field output factors for both square and circular fields were measured using nine different detectors. The corrected output factors (ratio of detector readings multiplied by the appropriate correction factor from the protocol) showed better consistency compared to the uncorrected output factors (ratio of detector readings only), with the relative standard deviation decreasing by approximately 1% after the application of the relevant correction factors. Comparisons relative to an arbitrarily chosen PTW 60019 microDiamond detector showed a reduction of maximal variation for the corrected values of approximately 3%. A full uncertainty budget was prepared to analyze the consistency of the output factors. Agreement within uncertainties between all detectors and field sizes was found, except for the 15 mm circular field. The results of this study show that the application of IAEA TRS 483(1) when measuring small fields will improve the consistency of small field measurements when using multiple detectors contained within the protocol.