4.7 Article

Effects of Surface Area on All-Solid-Stated pH Sensor Based on Antimony Electrode

期刊

IEEE SENSORS JOURNAL
卷 20, 期 2, 页码 680-688

出版社

IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
DOI: 10.1109/JSEN.2019.2945997

关键词

Antimony electrode; electrode surface area; pH sensor; soilless culture

资金

  1. National Key Technology R&D Program of China [2014BAD08B04]
  2. Special Fund for Agro-scientific Research in the Public Interest of China [201203095]
  3. Innovation Project for Graduate Student Research of Jiangsu Province of China [KYLX15_1055]

向作者/读者索取更多资源

To realize the on-line and in situ pH detection in soilless culture, a new all-solid-stated pH sensor was presented. The pH sensor consisted of alumina ceramic (Al2O3) substrate, Sb/Sb-2 O-3 working electrode, Ag/AgCl reference electrode, Ag lead and pad. Sb thin film and Ag thin film were obtained by magnetron sputtering method, and then thermal annealing and Nafionmodification was conducted to improve the performance of the sensor. Moreover, the influence of surface area of Sb electrode on the performance of pH sensor was investigated with the area of 4 mm(2), 9 mm(2), 16 mm(2), 25 mm(2) and 36 mm(2), respectively. Through surface morphology analysis, energy spectrum analysis and electrochemical experiments, the pH sensor, which Sb electrode surface area is 16 mm(2), shows excellent comprehensive performance, with high sensitivity (67.3 mV/pH), short response time (less than 3.5 s), good stability (standard deviation less than 2 mV in 700 s period), low hysteresis (less than 1.63 mV) and low drift rate (less than 1.62 mV/h). The developed sensor can be applied in pH detection in soilless culture and is promising to popularize in industrial production.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据