An international multi-center serum protein electrophoresis accuracy and M-protein isotyping study. Part I: factors impacting limit of quantitation of serum protein electrophoresis

期刊

CLINICAL CHEMISTRY AND LABORATORY MEDICINE
卷 58, 期 4, 页码 533-546

出版社

WALTER DE GRUYTER GMBH
DOI: 10.1515/cclm-2019-1104

关键词

accuracy; immunofixation; immunosubtraction; limit of quantitation; monoclonal proteins; protein electrophoresis

资金

  1. Sebia, Inc.
  2. Department of Laboratory Medicine and Pathology, Mayo Clinic, Rochester, MN, USA

向作者/读者索取更多资源

Background Serum protein electrophoresis (SPEP) is used to quantify the serum monoclonal component or M-protein, for diagnosis and monitoring of monoclonal gammopathies. Significant imprecision and inaccuracy pose challenges in reporting small M-proteins. Using therapeutic monoclonal antibody-spiked sera and a pooled beta-migrating M-protein, we aimed to assess SPEP limitations and variability across 16 laboratories in three continents. Methods Sera with normal, hypo- or hypergammaglobulinemia were spiked with daratumumab, Dara (cathodal migrating), or elotuzumab, Elo (central-gamma migrating), with concentrations from 0.125 to 10 g/L (n = 62) along with a beta-migrating sample (n = 9). Provided with total protein (reverse biuret, Siemens), laboratories blindly analyzed samples according to their SPEP and immunofixation (IFE) or immunosubtraction (ISUB) standard operating procedures. Sixteen laboratories reported the perpendicular drop (PD) method of gating the M-protein, while 10 used tangent skimming (TS). A mean percent recovery range of 80%-120% was set as acceptable. The inter-laboratory %CV was calculated. Results Gamma globulin background, migration pattern and concentration all affect the precision and accuracy of quantifying M-proteins by SPEP. As the background increases, imprecision increases and accuracy decreases leading to overestimation of M-protein quantitation especially evident in hypergamma samples, and more prominent with PD. Cathodal migrating M-proteins were associated with less imprecision and higher accuracy compared to central-gamma migrating M-proteins, which is attributed to the increased gamma background contribution in M-proteins migrating in the middle of the gamma fraction. There is greater imprecision and loss of accuracy at lower M-protein concentrations. Conclusions This study suggests that quantifying exceedingly low concentrations of M-proteins, although possible, may not yield adequate accuracy and precision between laboratories.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据