An international multi-center serum protein electrophoresis accuracy and M-protein isotyping study. Part II: limit of detection and follow-up of patients with small M-proteins

期刊

CLINICAL CHEMISTRY AND LABORATORY MEDICINE
卷 58, 期 4, 页码 547-559

出版社

WALTER DE GRUYTER GMBH
DOI: 10.1515/cclm-2019-1105

关键词

accuracy; immunofixation; immunosubtraction; limit of detection; monoclonal proteins; precision; protein electrophoresis

资金

  1. Sebia, Inc.
  2. Department of Laboratory Medicine and Pathology, Mayo Clinic, Rochester, MN, USA

向作者/读者索取更多资源

Background Electrophoretic methods to detect, characterize and quantify M-proteins play an important role in the management of patients with monoclonal gammopathies (MGs). Significant uncertainty in the quantification and limit of detection (LOD) is documented when M-proteins are <10 g/L. Using spiked sera, we aimed to assess the variability in intact M-protein quantification and LOD across 16 laboratories. Methods Sera with normal, hypo- or hyper-gammaglobulinemia were spiked with daratumumab or elotuzumab, with concentrations from 0.125 to 10 g/L (n = 62) along with a beta-migrating sample (n = 9). Laboratories blindly analyzed samples according to their serum protein electrophoresis (SPEP)/isotyping standard operating procedures. LOD and intra-laboratory percent coefficient of variation (%CV) were calculated and further specified with regard to the method (gel/capillary electrophoresis [CZE]), gating strategy (perpendicular drop [PD]/tangent skimming [TS]), isotyping (immunofixation/immunosubtraction [ISUB]) and manufacturer (Helena/Sebia). Results All M-proteins >= 1 g/L were detected by SPEP. With isotyping the LOD was moderately more sensitive than with SPEP. The intensity of polyclonal background had the biggest negative impact on LOD. Independent of the method used, the intra-laboratory imprecision of M-protein quantification was small (mean CV = 5.0%). Low M-protein concentration and high polyclonal background had the strongest negative impact on intra-laboratory precision. All laboratories were able to follow trend of M-protein concentrations down to 1 g/L. Conclusions In this study, we describe a large variation in the reported LOD for both SPEP and isotyping; overall LOD is most affected by the polyclonal immunoglobulin background. Satisfactory intra-laboratory precision was demonstrated. This indicates that the quantification of small M-proteins to monitor patients over time is appropriate, when subsequent testing is performed within the same laboratory.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据