Evaluation of SD Bioline HIV/syphilis Duo rapid test kits in Nepal

期刊

BMC INFECTIOUS DISEASES
卷 16, 期 -, 页码 -

出版社

BIOMED CENTRAL LTD
DOI: 10.1186/s12879-016-1694-9

关键词

Duo test; Sensitivity; Specificity; Kappa coefficient

资金

  1. National Center for AIDS and STD Control (NCASC)
  2. Family Health Division (FHD)

向作者/读者索取更多资源

Background: Accurate and prompt diagnosis of HIV and syphilis simultaneously has reinforcing effect on their control program because of their prevalent co-infection. Availability of a simple user-friendly two-pronged and affordable detection tools brings down the cost of health care. They are important in the antenatal clinics, with added opportunity for intervention and prevention of mother to child transmission. In cooperation with rapid test kit manufacturers, SD Bioline, NPHL and NCASC, an evaluation of commercially available HIV/syphilis Duo rapid test kit (SD Bioline) to assess its performance and operational characteristics was done in the present study. Method: A prospective laboratory-based cross sectional study was conducted at a large Women's Hospital. Ten thousand pregnant women, visiting the Hospital for antenatal care or for delivery, were enrolled in study. Tests were performed by the SD Bioline HIV/Syphilis Duo kit as well as national algorithm for HIV and syphilis diagnosis which were considered gold standard. Sensitivity, Specificity, positive predictive value and negative predictive value along with kappa coefficient were calculated for the kit under evaluation. Result: The sensitivity, specificity, Negative predictive value and Positive predictive value of the kit for HIV diagnosis were 100 % (95 % CI 83.18-100 %, 99.96-100 %, 83.18-100 %, and 99.96-100 %, respectively). Kappa value was found to be 1.0. Out of total cases, results of 9985 (99.85 %) cases were concordant with National algorithm for syphilis diagnosis. Thirteen (0.13 %) cases were found false positive while two were false negative. The sensitivity of the kit for syphilis diagnosis was found to be 95.45 % (95 % CI 84.86-98.74 %) and specificity was 99.87 % (95 % CI; 99.78-99.92 %). Positive predictive value was 76.36 % (95 % CI; 63.65-85.63 %) and Negative predictive value was 99.89 % (95 % CI; 99.39-99.99 %). Kappa value was found to be 0.85. Conclusion: The performance characteristics of SD Bioline HIV/Syphilis duo kit were found almost concordant with the kits being used for HIV and Syphilis diagnosis separately. Its implementation in antenatal clinics/VCTs could be an added opportunity for simultaneous diagnosis of HIV and syphilis.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据